Skip to main content

Table 1 The specimens’ mean and standard deviations of the measured parameters during the flexural test. P-values comparisons between control and disinfection/sterilization groups

From: Effects of autoclaving and disinfection on 3D surgical guides using LCD technology for dental implant

Groups

Control (GC)

Mean ± Std Dev

Autoclave 121 °C (GA)

Mean ± Std Dev

p-value

Autoclave 134 °C (GB)

Mean ± Std Dev

p-value

Disinfected

(GL)

Mean ± Std Dev

p-value

Flexural Strength (MPa)

58,9 ± 4,12

70,64 ± 7,71

0.258

74,24 ± 16,67

0.113

79,94 ± 11,55

0.042

Flexural Strain %

8,95 ± 0,35

3,95 ± 0,72

0.003

8,03 ± 0,78

0.308

7,35 ± 0,31

0.089

Flexural Modulus of Elasticity (Mpa)

575,11 ± 114,64

1804,05 ± 144

0.002

1031,61 ± 171,88

0.174

1082,26 ± 114,41

0.089