Skip to main content

Table 2 The specimen’s mean and standard deviations of the measured parameters during the tensile test. P-values comparisons between control and disinfection/sterilization groups

From: Effects of autoclaving and disinfection on 3D surgical guides using LCD technology for dental implant

Groups

Control (GC)

Mean ± Std Dev

Autoclave 121 °C (GA)

Mean ± Std Dev

p-value

Autoclave 134 °C (GB)

Mean ± Std Dev

p-value

Disinfected

(GL)

Mean ± Std Dev

p-value

Tensile Strength (MPa)

36,07 ± 4,45

8,47 ± 0,79

0.013

11,55 ± 1,31

0.141

36,50 ± 1,30

0.91

Tensile Strain %

7,23 ± 2,31

5,48 ± 0,81

0.46

3,37 ± 0,42

0.008

4,74 ± 0,17

0.257

Tensile Modulus of Elasticity (Mpa)

522,72 ± 123,83

157,37 ± 30,74

0.042

344,54 ± 39,87

0.308

770,36 ± 19,61

0.308