From: Material characterization and selection for 3D-printed spine models
| Moment in Y-Direction (My) | ||
|---|---|---|
| Material Combination | Maximum Peak (N-cm) | vs. Vertebrae (N) |
| Non-UV | 4.9 | 0.8 |
| UV | 3.2 | −0.9 |
| 10% Tough + 90% Clear | 3 | −1.1 |
| 100% Tough | 3.5 | − 0.6 |
| 0.125% Kevlar + 100% Clear | 8.2 | 4.1 |
| 10% Castable + 90% Clear | 11.8 | 7.7 |
| 5% Gypsum + 100% Clear | 6 | 1.9 |
| 15% Gypsum + 100% Clear | 6 | 1.9 |
| 30% Gypsum + 100% Clear | 15 | 10.9 |
| 10% Flexible + 90% Clear | 7.6 | 3.5 |
| 5% Flexible + 5% Tough + 90% Clear | 5 | 0.9 |
| 100% Dental | 11.9 | 7.8 |
| PLA Unpolished | 4.2 | 0.1 |