Skip to main content

Table 6 Summary of Dose Errors

From: Dose calibration of Gafchromic EBT3 film for Ir-192 brachytherapy source using 3D-printed PLA and ABS plastics

Dose Level

Material

Color Channel

Error (%)

(%Uncertainty)

High

Water

Red

0.8

(6.86%)

 

Green

0.3

 

ABS

Red

−6.9

  

Green

−5.2

 

PLA

Red

−7.8

  

Green

−5.7

Low

Water

Red

4.8

(8.36%)

 

Green

3.2

 

ABS

Red

−5.8

  

Green

−6.9

 

PLA

Red

−10.6

  

Green

−15.8

  1. Table 6 Legend: Summary of error values for red and green channels in Water, ABS, and PLA at high and low dose levels with uncertainties