Skip to main content

Table 1 Physical measurements of cadaveric hearts and 3D printed models

From: Methods for verification of 3D printed anatomic model accuracy using cardiac models as an example

Cardiac Valve

Feature

Model #

3D model measurement/ cadaveric model measurement in mm (% difference)

± Standard Deviation across 4 observers

Observer 1

Observer 2

Observer 3

Observer 4

3D Model (mm)

Cadaveric heart (mm)

% Difference

Aortic

STJ perimeter

1

105.0/104.0 (101.0)

106.0/103.0 (102.9)

106.0/101.0 (105.0)

97.0/106.0 (92.0)

4.4

2.1

5.7

Aortic

STJ perimeter

2

103.0/98.0 (105.1)

105.0/100.0 (105)

87.0/89.0 (97.8)

107.0/97.0 (110.3)

9.2

4.8

5.1

Aortic

STJ perimeter

3

101.0/106.0 (95.3)

103.0/102.0 (101.0)

100.0/103.0 (97.1)

106.0/111.0 (95.5)

2.6

4.0

2.6

Aortic

Max Diameter

1

34.4/32.8 (104.9)

34.6/34.5 (100.3)

35.1/34.1 (102.9)

32.1/33.2 (96.7)

1.3

0.8

3.6

Aortic

Max Diameter

2

33.3/32.5 (102.5)

34.5/32.4 (106.5)

29.6/28.1 (105.3)

34.4/34.4 (100.0)

2.3

2.7

2.9

Aortic

Max Diameter

3

32.5/31.1 (104.5)

32.3/31.0 (104.2)

29.9/32.7 (91.4)

35.3/35.6 (99.0)

2.2

2.2

6.1

Aortic

Min diameter

1

29.6/29.7 (99.6)

30.6/29.9 (102.3)

29.4/31.0 (94.8)

27.3/29.2 (93.7)

1.4

0.8

4.1

Aortic

Min diameter

2

27.6/25.6 (107.8)

27.6/26.4 (104.5)

23.9/23.4 (102.1)

27.4/27.4 (100.0)

1.8

1.7

3.4

Aortic

Min diameter

3

28.0/27.8 (100.7)

28.1/28.4 (98.9)

31.8/26.6 (119.5)

29.6/30.2 (98.1)

1.8

1.5

10.2

Mitral

Annulus perimeter

4

120.0/125.0 (96.0)

116.0/127.0 (91.3)

121.0/130.0 (93.1)

124.0/126.0 (98.4)

3.3

2.2

3.1

Mitral

Annulus perimeter

5

122.0/125.0 (97.6)

112.0/119.0 (94.1)

110.0/110.0 (100.0)

101.0/121.0 (83.5)

8.6

6.3

7.3

Mitral

Max Diameter

4

43.5/45.2 (96.2)

43.5/43.2 (100.7)

43.3/45.3 (95.6)

37.9/37.9 (100.0)

2.8

3.5

2.6

Mitral

Max Diameter

5

40.2/43.6 (92.2)

40.0/40.9 (97.8)

47.5/45.3 (104.9)

40.4/35.3 (114.7)

3.6

4.4

9.7

Mitral

Min diameter

4

26.5/28.3 (93.6)

25.1/24.9 (100.8)

25.3/27.5 (92.0)

27.3/27.3 (100)

1.0

1.5

4.4

Mitral

Min diameter

5

23.9/24.5 (97.6)

24.0/28.3 (84.8)

25.6/30.3 (84.5)

23.0/29.5 (77.9)

1.1

2.6

8.2

  1. Measurements were made of aortic valves on 3 hearts and mitral valves on 2 hearts by 4 observers. Aortic valve features measured included the perimeter of the sinotubular junction (STJ), which is important in pre-procedural planning for transcatheter aortic valve replacement (TAVR), and the maximum and minimum valve diameter. Mitral valve features included the perimeter of the mitral annulus, as well as the maximum and minimum diameter of the mitral valve. 3D model and cadaveric heart measurements are presented in millimeters (mm), along with the percentage difference between the 3D model and accompanying cadaveric heart measurements, for each observer. Percentage differences exceeding 100% indicate that the model measurement was larger than the cadaveric heart measurement. The standard deviation for measurements across the 4 observers is listed in the right-hand columns for 3D model feature measurement, cadaveric heart measurement, and % difference between the 3D model and cadaveric heart measurements