Fig. 7 | 3D Printing in Medicine

Fig. 7

From: 3D-printed saw guides for lower arm osteotomy, a comparison between a synthetic CT and CT-based workflow

Fig. 7

Distance mapping of surface model comparison. sCT (a) and CT (b) surface models compared to the micro-CT surface model of proximal radius. The color bar indicates differences (mm) between the micro-CT and the sCT or CT within a − 1 and 1 mm range. A positive value indicates a larger sCT or CT model than the micro-CT model. c Exhibits an inconsistency in the sCT surface model compared to the micro-CT surface model of a distal radius. The corresponding sCT saw guide was a placement outlier during the saw guide placement study. d Shows the CT surface model compared to the micro-CT surface model of the same distal radius