Table 1 Average (Standard deviation (SD)) absolute translation and rotation errors of the CT- and sCT-based saw guides placed by the six observers. The p-value was calculated with a rANOVA on the between subjects differences of each parameter, with p < 0.025 being significantly different

From: 3D-printed saw guides for lower arm osteotomy, a comparison between a synthetic CT and CT-based workflow

Saw guide type

Translation: mm (SD)

Rotation: degrees (SD)

 

∆x

∆y

∆z

∆T average

∆T max diff

ϕx

ϕy

ϕz

∆R average

∆R max diff

CT-based

0.8 (±1.1)

0.4 (±0.6)

2.1 (±2.3)

2.4 (±2.4)

4.5 (±3.4)

0.6 (±0.7)

0.5 (±0.6)

3.5 (±4.8)

3.8 (±4.8)

6.9 (±6.8)

sCT-based

1.0 (±1.3)

0.5 (±0.6)

2.3 (±2.4)

2.8 (±2.5)

4.5 (±2.8)

0.7 (±0.6)

0.8 (±0.7)

4.6 (±6.0)

4.9 (±6.0)

7.0 (±6.8)

p-value

.892

.687

.752

.284

.245

.167

.227

.216