Table. 2 Estimated differences between the original design and the 3D-printed pieces before (Comparison1) and after the sterilization process (Comparison3)

From: Does vaporized hydrogen peroxide sterilization affect the geometrical properties of anatomic models and guides 3D printed from computed tomography images?

  Measured points (n) Mean difference Comparison1a SDa Lower 95%CIa Upper 95%CIa Mean difference Comparison3a SDa Lower 95%CIa Upper 95%CIa
Model Number
 1 168098 −0,019 0,467 −0,021 − 0,017 0,017 0,529 0,015 0,020
 2 144685 −0,037 0,398 −0,039 −0,035 − 0,023 0,452 − 0,025 −0,020
 3 242098 −0,168 0,469 −0,170 −0,166 − 0,165 0,516 − 0,167 −0,163
 4 114841 −0,072 0,529 −0,075 −0,069 − 0,071 0,623 − 0,073 −0,066
 5 100099 0,021 0,570 0,017 0,024 0,074 0,719 0,069 0,078
 6 128057 −0,111 0,506 − 0,114 −0,108 − 0,075 0,630 − 0,078 −0,071
 7 215781 −0,139 0,620 −0,141 −0,136 − 0,179 0,737 − 0,182 −0,176
 8 171893 −0,134 0,643 −0,138 −0,131 − 0,097 0,723 − 0,100 −0,094
Total 1285552 −0,095 0,536 −0,096 −0,094 − 0,082 0,626 − 0,083 -0,081
Guide Number
 1 47785 0,115 0,189 0,114 0,117 0,119 0,198 0,117 0,121
 2 116214 0,091 0,207 0,090 0,092 0,100 0,194 0,099 0,101
 3 192570 0,162 0,210 0,161 0,163 0,116 0,193 0,115 0,116
 4 73695 0,100 0,184 0,098 0,101 0,083 0,174 0,082 0,084
 5 138000 0,244 0,355 0,244 0,245 0,238 0,259 0,237 0,240
 6 108202 0,118 0,188 0,117 0,119 0,092 0,175 0,091 0,093
 7 69266 0,071 0,175 0,070 0,073 0,098 0,140 0,097 0,099
 8 23113 0,093 0,195 0,091 0,096 0,071 0,170 0,069 0,073
Total 768845 0,141 0,240 0,140 0,141 0,126 0,205 0,126 0,127
  1. aValues are expressed in millimeters. SD Standard deviation, CI Confidence interval