From: Design, inverted vat photopolymerization 3D printing, and initial characterization of a miniature force sensor for localized in vivo tissue measurements
Run
Applied
ΔRSG
Experimental
Theoretical
Error
Order
Load (mN)
(Ω)
Strain (με)
(%)
3
28
-0.11
10.81
11.14
3.13
10
38
-0.15
14.73
14.65
2.12
1
214
-0.87
85.46
86.28
1.89
2
313
-1.28
125.73
125.91
1.40
4
509
-2.06
202.35
204.55
1.18