Skip to main content

Table 1 Mean, SD, range and statistical analysis of diameter, thickness, and articular depth

From: Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study

 

3Dmodel-CTA

3Dmodel-CT

3Dmodel-CIRM

Friedman test (p-value)

Pairwise comparisons

(Bonferroni correction)

mean ± SD range (mm)

mean ± SD range (mm)

mean ± SD range (mm)

3Dmodel-CTA vs. 3Dmodel-CT (p-value)

3Dmodel-CTA vs. 3Dmodel-CIRM (p-value)

3Dmodel-CIRM vs. 3Dmodel-CT (p-value)

Diameter along coronal plane

21.85±0.99

20.68–23.55

20.86±0.94

19.94–22.81

21.66±0.91

20.64–23.39

0.002*

0.003*

1.000

0.018

Diameter along sagittal plane

22.39±1.52

20.45–24.13

21.51±1.43

19.34–23.62

22.07±1.49

19.83–24.03

0.001*

0.001*

0.401

0.073

Thickness at medial rim

10.98±1.23

9.24–12.32

9.91±0.96

8.55–11.57

10.75±1.05

9.26–12.30

0.002*

0.003*

1.000

0.018

Thickness at lateral rim

8.46±0.9

6.74–9.61

7.5±0.97

5.91–9.18

8.35±0.94

6.61–9.82

0.002*

0.008*

1.000

0.008*

Articular disc depth

2.03±0.20

1.71–2.38

1.63±0.19

1.35–1.91

1.75±0.12

1.58–1.94

0.002*

0.001*

0.037

0.952

  1. *significant at p-value < 0.01
  2. 3Dmodel-CTA 3D model from CT arthrography, 3Dmodel-CT 3D model from CT scan, 3Dmodel-CIRM 3D model from cartilage-reproducing image reconstruction method