From: Use of patient specific 3D printed neurovascular phantoms to simulate mechanical thrombectomy
Experimental Variable | Statistical Test | Mean (Successful) | Mean (Unsuccessful) | P-Value |
---|---|---|---|---|
Clot Angulation | Student T-test | 118.10 ± 19.44° | 153.46 ± 17.37° | 0.016 |
Clot Length | Student T-test | 16.76 ± 2.64 mm | 15.38 ± 3.68 mm | 0.557 |
Clot Composition | Chi-Squared Test | Â N/A | Â N/A | 0.115 |
Clot Location in Device | Chi-Squared Test | Â N/A | Â N/A | 0.196 |
Clot Location in Vasculature | Chi-Squared Test | Â N/A | Â N/A | 0.579 |