Skip to main content

Table 3 Descriptive outcome measures for the deviation of 3D model dimensions (mm) for different AM technologies and (part) orientations

From: Dimensional accuracy and precision and surgeon perception of additively manufactured bone models: effect of manufacturing technology and part orientation

Technology

Orientation

Mean

SD

RMSE

MIN

MAX

Median

IQR

DLP

diagonal

0.12

0.09

0.14

-0.76

0.69

0.12

0.09

DLP

horizontal

0.13

0.16

0.20

-0.96

0.75

0.14

0.19

DLP

vertical

0.12

0.05

0.13

-0.28

0.84

0.12

0.04

FDM

diagonal

0.00

0.08

0.08

-0.56

0.71

0.00

0.09

FDM

horizontal

0.07

0.16

0.18

-0.94

1.04

0.06

0.17

FDM

vertical

0.00

0.07

0.07

-0.68

0.73

0.00

0.07

PJ

diagonal

0.08

0.05

0.09

-0.19

0.58

0.07

0.06

PJ

horizontal

0.06

0.05

0.08

-0.24

0.30

0.06

0.07

PJ

vertical

0.11

0.07

0.13

-0.79

0.51

0.11

0.08

  1. Abbreviations: SD: standard deviation, RMSE: root mean squared error, MIN: minimum, MAX: maximum, IQR: interquartile range