Table 4 Mechanical property evaluations after worst-case exposure to vaporized hydrogen peroxide

From: Low temperature vaporized hydrogen peroxide sterilization of 3D printed devices

Test Name

ASTM

Material / % change

Formlabs BioMed Amber

Formlabs

BioMed Clear

Stratasys MED610

Stratasys MED615

Stratasys MED620

Tensile Strength

D638

-8.5% (V)

-3.4% (H)

-5.0% (V)

-5.3% (H)

-1.8% (V)

-14.7% (H)

4.2% (V)

-16.6% (H)

4.2% (V)

-16.6% (H)

Flexural Strength

D790

-9.0% (V)

-8.2% (H)

-6.8% (V)

-3.6% (H)

-17.2% (V)

-14.4% (H)

-21.6% (V)

-16.8% (H)

-21.6% (V)

-16.8% (H)

Compressive Strength

D695

-4.7% (V)*

-8.6% (H)*

29.2% (V)*

9.7% (H)*

-0.8% (V)

15.4% (H)

15.7% (V)

8.1% (H)*

15.7% (V)

8.1% (H)*

Izod notched impact

D256

-9.8% (V)

21.7% (H)

7.2% (V)

32.1% (H)

9.7% (V)

-4.6 (H)

19.0% (V)

24.3% (H)

19.0% (V)

24.3% (H)

Shore Hardness

D2240

1.0%

0.0%

1.2%

-0.7% *

-0.7% *

  1. Testing was conducted by Westmoreland Mechanical Testing & Research, Inc., an A2la ISO 17,025 accredited and NADCAP accredited laboratory
  2. * The exposed sample result is Not Statistically Significant compared to the unexposed control