Skip to main content

Table 7 Van Dal evaluation of formlabs BioMed clear

From: Low temperature vaporized hydrogen peroxide sterilization of 3D printed devices

Test Name

Steam Sterilization 134 °C, 3.5 min

This study

0° Print, % change

45° Print, % change

Tensile Strength

-5.8

1.4

-5.3% (H)

Flexural Strength

31

30

-3.6% (H)

Notched Impact Strength

-17

-21

32.1% (H)