From: Low temperature vaporized hydrogen peroxide sterilization of 3D printed devices
Test Name | Steam Sterilization 134 °C, 3.5 min | This study | |
---|---|---|---|
0° Print, % change | 45° Print, % change | ||
Tensile Strength | -5.8 | 1.4 | -5.3% (H) |
Flexural Strength | 31 | 30 | -3.6% (H) |
Notched Impact Strength | -17 | -21 | 32.1% (H) |