Table 2 Percentage of successful resections within accepted surgical error margin

From: Fluoroscopically calibrated 3D-printed patient-specific instruments improve the accuracy of osteotomy during bone tumor resection adjacent to joints

Surgical error margin

2 mm

3 mm

4 mm

5 mm

6 mm

7 mm

8 mm

9 mm

10 mm

FCPSI

33.3%

66.7%

100%

100%

100%

100%

100%

100%

100%

CPSI

0.0%

16.7%

33.3%

66.7%

83.3%

83.3%

100%

100%

100%

Freehand

0.0%

0.0%

0.0%

0.0%

33.3%

66.7%

100%

100%

100%